femometer IR109 Infrared Thermometer User Manual

Home » Femometer » femometer IR109 Infrared Thermometer User Manual

femometer logo

femometer IR109 Infrared Thermometer

femometer IR109 Infrared Thermometer

Introduction

Thank you for purchasing the ear infrared thermometer. In order to use it correctly, please read this instruction carefully before using the product, and operate and use it in accordance with the requirements of the manual. Be sure to read and follow the “Warnings and Precautions” before using this product.  Please keep the manual with the product so that can refer to it at any time.

Package Contents

No. Name Quantity
1 Infrared Thermometer 1
2 Battery (AA, optional) 2
3 User Manual 1
4 Pouch 1
5 Disposable Hygiene Cap 20 (supplied)

+1 (installed)

Warnings and Precautions

  • Please consult your physician if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc. in spite of any colour of the background light or absence of fever. The operating ambient temperature range for this thermometer is 50°F(10°C ) -104°F(40°C). Do not expose the thermometer to temperature extremes (below -4°F(-20°C) / or over 131°F(55°C )) or excessive humidity (>95% RH).
  • Please consult your physician if the thermometer shows elevated temperature (yellow or red background lights).
  • This thermometer must only be used with Disposable Hygiene Caps. To avoid inaccurate measurements always use this thermometer with a new, clean Hygiene cap attached.
  • If the thermometer is accidentally used without a Hygiene cap attached, clean the lens (see “Care and cleaning” section). Keep Hygiene caps out of reach of children.
  • This thermometer is intended for household use only. This product is not intended to diagnose any disease, but is a useful screening tool for temperature. Use of this thermometer is not intended as a substitute for consultation with your doctor.
  • Do not modify this equipment without authorization of the manufacturer.
  • Parents/guardians should call the pediatrician upon noticing any unusual sign(s) or symptom(s). For example, a child who exhibits irritability, vomiting, diarrhea, dehydration, seizure, changes in appetite or activity, even in the absence of fever, or who exhibits a low temperature, may still need to receive medical attention.
  • Children who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to determine the severity of their illness.
  • Temperature elevation as indicated by Age Precision may signal a serious illness, especially in adults who are old, frail, have a weakened immune system, or neonates and infants. Please seek professional advice immediately when there is a temperature elevation and if you are taking temperature on:
  • neonates and infants under 3 months (consult your doctor immediately if the temperature exceeds 99.4 °F or 37.4 °C)
  • patients over 60 years of age.
  • patients having diabetes mellitus or a weakened immune system (e.g. HIV positive, cancer chemotherapy, chronic steroid treatment, splenectomy)
  • patients who are bedridden (e.g. nursing home patient, stroke, chronic illness, recovering from surgery)
  • a transplant patient (e.g. liver, heart, lung, kidney)Fever may be blunted or
    even absent in elderly patients.
  • This thermometer contains small parts that can be swallowed or produce a choking hazard to children. Always keep the thermometer out of children’s reach.

Product Description

This ear thermometer is intended to be used by home users for intermitient measurement of the body temperature of pa?ents having ages ranging from normal weight (full term) new-born to geriatric adult.

  • This thermometer is not intended for clinical use in a professional environment and are for home usage only.
  • The Age Precision feature is not intended for pre-term babies or small-forgestational age babies.
  • This thermometer is not intended to interpret hypothermic temperatures.
  • Do not allow children under 12 to take their temperatures unattended.

Product Features

  • Three ages mode(Neonate/Baby/Adult modes)
  • 30 groups of measurement data memory function
  • Mute/Un-mute function
  • One-touch Pop-up ear Hygiene Cap
  • Display Time
  • Fever warning function
  • Fahrenheit/Celsius unit conversion
  • With 21 Disposable Hygiene Cap
  • Recognize the hygiene cap status automatically
  • Automatic shutdown function, low power consumption

Product Structurefemometer IR109 Infrared Thermometer 1

  1. Cap Ejector
  2. LCD Display Screen
  3. Start Button
  4. Age Button
  5. Memory Button
  6. Probe Cover
  7. Hygiene Cap
  8. Probe Tip
  9. Rating Label
  10. Battery Cover

Display Descriptionfemometer IR109 Infrared Thermometer 2

  1. Temperature Value
  2. Time Display
  3. Ear Temperature Mode
  4. Hygiene Cap Icon
  5. Fahrenheit/Celsius Degrees
  6. Memory Mode
  7. Mute/Un-mute Icon
  8. Battery Icon
  9. Adult Mode (36 Months-adult)
  10. Baby Mode (3-36 Months)
  11. Neonate Mode ( 0-3 Months)

How to Use Your Thermometer

When using the thermometer for the first time, please install the batteries.

  1. Wear Disposable Hygiene Cap Correctly
    First remove the head cover, put on the hygiene cap tightly , then press the  button to turn on the thermometer.femometer IR109 Infrared Thermometer 3
    Note: If you don’t wear the hygiene cap tightly, the Hygiene cap indicator will blink, you can’t measure the temperature.
  2. Age Mode Convertion
    When turned on, press “AGE” button to toggle through each age group.

    1.  Neonate Mode (0-3 months baby) When the screen shows “ ”, it turns to neonate mode;
    2. Baby Mode (3-36 months baby)
      When the screen shows “ ”, it turns to baby mode;
    3. Adult Mode (over 36 months age)
      When the screen shows “ ” , it turns to adult mode.
      Note: You must select an age in order to take a measurement. The default mode is Adult Mode.
  3. Time Setting
    When turn on the thermometer, press the “MEM” button for 3 seconds, release the button when the minute digit is flashing. Press the “MEM” button to switch H/M digit when it’s flashing; Press the “AGE” button to set the hour time; The time setting will be finished when press the Measure button or without any operation for 6 seconds.

    femometer IR109 Infrared Thermometer 4

  4. Measure the Temperature of Your Ears
    Fit the probe snuggly into the ear canal and direct towards opposite temple. Then push and release the Measure button “ ”.
    NOTE: Please be aware of the factors that influence the accuracy as described in the section “Temperature Taking Tips” and “Warnings and Precautions”.femometer IR109 Infrared Thermometer 5
    Always clean the lens before use. Always make sure the user and the thermometer are in the same room for at least 30 minutes before the measurement.
  5. Take Room/Object Temperature
    1. When power off, press the MEM button for 3 seconds . The screen will display °F/°C, then release the button, you will hear a beep, and the screen will be into a full display, then it will complete a temperature test after 1 second.
    2. It will be the same as the right picture showed. When the °F/°C is flashing, then it’s ready for testing.
    3. Press the measure button, test the temperature. * When taking Room/Object Temperature,no need to wear Hygiene Cap for measuring.

      femometer IR109 Infrared Thermometer 6

  6. After the Measurement
    Once the reading has been completed, remove the thermometer away from the ear and observe temperature. After each measurement, you can enter the recall mode and query earlier temperature readings. Do not hold the thermometer for a long time, because it is sensitive to the ambient temperature. After each measurement, clean the temperature probe with a soft cloth, and put the thermometer in a dry and well-ventilated place. It is recommended to wait for 10 seconds between each measurement. It is dangerous to make a self-diagnosis or self-treatment based on the obtained measurement results. For such purposes, please consult a doctor.
  7. Read Your Temperature
    “T” indicates a temperatrue reading.femometer IR109 Infrared Thermometer 7
    NOTE: For the next measurement, press the button to remove and discard used hygiene cap, and then put on a new hygiene cap.

    femometer IR109 Infrared Thermometer 8

  8. Switching between Mute and Un-mute
    When the thermometer is turned on, press the “ AGE” button for 3 seconds to turn the sound on or off.
  9. Check 30 Sets of Memory Data
    When the thermometer is turned on , short press the “Mem” button to go to the memory mode, and press “ Mem” button again to check the 30 sets of memories one by one. If no value, it will display “—M”.
  10. °F/°C Conversion
    Open the battery cover, use the toggle switch to change the °F/°C.
  11. To Turn Off
    The unit will shut down automatically after “30±1 seconds” without using. Keep pressing the Measure button for more than 6 seconds, it will shut down.femometer IR109 Infrared Thermometer 9
    Caution: 

    1. All memory records will loss when uninstall or reinstall the batteries.
    2. All settings will come to default when uninstall the batteries. If need adjust the settings, please turn on and make the new settings.
  12. Replace the Battery
    Press the battery button on the back bottom of thermometer to open the battery compartment. Put two AA batteries correctly into the compartment.
    Remove the batteries, if the thermometer will not be used for more than two months.

Disposal
For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose of the unit at a suitable local collection or recycling point.

Observe the local regulations for material disposal

  • Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal. are required by law to recycle used batteries.
  • You can recycle your old batteries at public collection points in your community or wherever batteries of the relevant type are sold.
  • Pb = Battery contains lead, Cd = Battery contains cadmium,
  • Hg = Battery contains mercury.
  • The batteries in this device do not contain any pollutants.

Temperature Taking Tips

  1. It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. Record readings twice a day (early morning and late afternoon). Take the average of the two temperatures to calculate normal equivalent temperature.
  2. Holding the thermometer for too long in the hand before taking a measurement can cause the device to warm up. This means the measurement could be incorrect.
  3. Patients and the thermometer should stay in steady-state room condition for at least 30 minutes.
  4. In the following situations it is recommended that 3-5 temperatures in the same location be taken and the highest one taken as the reading:
    • Newborn infants in the first 100 days.
    • Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
    • When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the instrument and obtains consistent readings.

Care and Cleaning

  1. Cleaning and user maintenance must not be performed by children unless supervised.
  2. Cleaning the device use an alcohol swab or cotton swab moistened with 75% alcohol to clean the thermometer case and the measuring probe.femometer IR109 Infrared Thermometer 10
  3. After the alcohol has completely dried out, you can take a new measurement (minimum of 10 minutes).
  4. Ensure that no water enters the device. If this happens, only use the device again once it has fully dried out.
  5. Never use abrasive cleaning agents, thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids.
  6. Take care not to scratch the surface of the LCD screen.
  7. Remove the batteries from the device before cleaning.
  8. Clean the device after use with a soft, slightly damp cloth. If they are very dirty, you can also moisten the cloth with a mild soapy solution.

Error and Troubleshootingfemometer IR109 Infrared Thermometer 11 femometer IR109 Infrared Thermometer 12

Specifications:

Product Name Infrared Thermometer
Power Supply DC1.5V×2
Displayed Temperature Range 89.6-109.2°F (32-42.9°C)
Accuracy (Laboratory) Maximum Laboratory Error ±0.4°F (±0.2°C)
Display Resolution 0.1°F/0.1°C
Automatic Shutdown 30s±1s
Memory 30 groups of measured temperature.
Battery 2*AA, can be used for more than 3000 times
Weight & Dimension 76.5g (without battery), 35*156*67mm
 

Operating Conditions

Operating environment: 50°F-104°F (10 °C-40°C),

≤85%RH

Atmospheric pressure: 70 – 106kPa

 

Storage

Temperature: -4°F~131°F (-20 °C~55 °C)

Relative Humidity: ≤95% moisture condensation

Atmospheric pressure: 70 – 106kPa

Service Life 5 Years

SYMBOLSfemometer IR109 Infrared Thermometer 13 femometer IR109 Infrared Thermometer 14

EMC Information

IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product. The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.

  • Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
  • Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  • Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30cm to any part of the Infrared Thermometer (FC-IR202), including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • If any: A list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of Clause7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g.shieldedcable, load impedance) or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE REFERENCE).
  • If any: The performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the
  • OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFORMANCE” need not be used).

Technical description:

  1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
  2. Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.

Table 1

Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic emissions IEC 61000-3-2 N/A
Voltage fluctuations/ flicker emissions

IEC 61000-3-3

 

N/A

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity Test IEC 60601-1-2 Test level Compliance level
Electrostatic discharge(ESD) IEC 61000-4-2 ±8 kV contact ±2 kV,

±4 kV, ±8 kV, ±15 kV air

±8 kV contact ±2 kV,

±4 kV, ±8 kV, ±15 kV air

 

Electrical fast transient

/burst IEC61000-4-4

Power supply lines: ±2 KV input/output lines: ±1 kV

lO0kHz repetition frequency

 

N/A

Surge IEC61000-4-5 line (s) to line (s): ±1 KV

line (s) to earth: ±2 KV

 

N/A

Voltage dips, short

interruptions and voltage variations on power supply input lines IEC61000-4-11

0% 0.5 cycle At 0″, 45: 90′, 135′,

180′, 225′, 270′ and 315′ 0% 1

cycle And 70% 25/30 cycles Single phase: at 0 0% 250 cyde

 

 

N/A

Power frequency magnetic

field IEC61000-4-8

30A/m

50Hz/60Hz

30A/m

50Hz/60Hz

Conduced RF IEC61000-4-6 150KHz to 80MHz: 3Vms 6Vrms (in ISM and amateur radio bands)

80%AM at 1 kHz

 

N/A

Radiated RF IEC61000-4-3 l0V/m, 80MHz-2,7GHz

80%AM at 1 kHz

l0V/m, 80MHz-2,7GHz

80%AM at 1 kHz

NOTE UT is the a.c. mians voltage prior to application of the test level.

Table 3

Guidance and manufacturer’s declaration-electromagnetic immunity
  Test Frequency (MHZ) Band

(MHZ)

 

Seivice

 

Modulation

Modulat-

ion (W)

Distance

(M)

IMMUNllY

TEST LEVEL

(V/m)

        Pulse      
Radiated 385 380-390 TETRA400 modulation 1.8 0.3 27
RF       18Hz      
IEC61000

-4-3 (Test specifica-

 

450

 

430-470

GMRS460, FRS460 FM ±5kHz

deviation

1 kHz sine

 

2

 

0.3

 

28

tions for
710   LTEBand 13, 17 Pulse modulation 217Hz      
ENCLO-

SURE PORT

704-787 0.2 0.3 9
745
780
IMMUNITY        
810   GSM800/900,TETRA Pulse      
to RF

wireless

800-960 800,iDEN 820, CDMA

850, LTE Band 5

modulation 18Hz 0.2 0.3 28
870
930
communi
1720   GSM 1800; CDMA Pulse      
cations equip- ment) 170-0 1990 1900; GSM 1900; DECT; LTE Band

1,3,4,25; UMTS

modulation 217HZ 2 0.3 28
1845
1970
    Bluetooth WLAN,802.11 Pulse      
  2450 2400-2570 b/g/n, RFID 2450, modulation 2 0.3 28
      LTE Band A 217HZ      
  5240 5100-5800 WLAN

802.11a/n

Pulse modulation

217HZ

 

0.2

 

0.3

 

9

5500
5785

Warranty and After-Sale Service

  • The device is under warranty for lifetime unlimited from the date of purchase.
  • The batteries, the packaging, and any damage caused by improper use are not covered by the warranty.
  • Excluding the following user-caused failures:
  1. Failure resulting from unauthorized disassembly and modification.
  2. Failure resulting from an unexpected dropping during application or transportation.
  3. Failure resulting from not following the instructions in the operating manual.

Leave a Comment