Introduction
This manual is an outline of the AD-8552EIP and the instructions for setting up and installing the equipment. Refer to the A&D website for more information on the compatible weighing devices and communication protocols.(https://www.aandd.jp/)
Features
The AD-8552EIP converts RS-232C communications of the weighing device into EtherNet/IP communications (EtherNet/IP converter).
- By using EtherNet / IP, the balance can be controlled directly from the network. In principle, the AD-8552EIP can connect up to 253 units to one PLC. (In actually, the maximum number of connected devices varies depending on the PLC performance and communication frequency.)
- The weighing value can be reset to zero (re-zero) by operating from the PLC.
- The hook on the back of this device allows one-touch DIN rail mounting.
- When connected to a AD-4212C, etc., response speed can be changed, Sensitivity adjustment can be done with an external weights, and power can be supplied to the measuring instrument. Please check A&D website for compatible measuring instruments.
Cautions
Before use, confirm the following items for safe operation.
- This device is a precision instrument. Please handle with care.
- Avoid vibration, shock, extremely high temperature and humidity, direct sunlight, dust, splashing water, air containing salt or corrosive gases, and places where flammable gases are present.
- The operating temperature is 0°C to +50°C (32°F to 122°F).
- Please ground the module.
- The power supply is DC24V. If it contains a momentary power failure or noise components, that may cause a malfunction. Use a stable power supply. Please avoid sharing with the power line.
- Do not share the earth ground line and power line with other electrical power equipment.
- Do not turn on the converter until installation is complete. The convertor is not equipped with a switch to turn it off.
- To prevent foreign matter from entering this device, do not remove the prevent overheating, be sure to remove the protective cover before turning on the power after installation and wiring.
Specification
4.Specification | |
4.1.Specification | |
Voltage requirement | DC 24 V +10%、-15% |
Power requirement | 9W Max. (When power is supplied to 4 units such as an AD-4212C) |
Communication
interface |
EtherNet/IP ×2 (For connection to control equipment)
RS-232C× 4 (For connection with a weighing instrument) |
Operating conditions | 0℃ to +50℃, Max 85 %RH (no condensation) |
External dimensions | 105 (W)×112 (H)×103 (D) mm |
Accessory | Simplified instruction manual |
Mass | Approximately 440 g |
4.2.Communication specification (EtherNet/IP) | |
Communication standard | EtherNet/IP (CT18 compliant) |
Vender ID | 188 |
Device type | 43(0x2B):Generic Device |
Transmission speed | 10/100Mbps (automatic negotiation) |
Communication method | Full duplex / half duplex (automatic negotiation) |
4.3.Communication specification (RS-232C) | |
Baud rate | 2400, 9600, 19.2k, 38.4k (Set by SW-3, 4.) |
Data bit length | 7 bit fixed |
Parity | EVEN fixed |
Terminator | <CR><LF> fixed |
Front and Rear Panel

Connections
Power supply
Connections example
- When connecting and removing the cables, push the buttons with a driver etc.
- We recommend use of rod type crimp terminals for the tips of cables.
Shields of the RS-232C connector are connected internally to the FG of the power connector.
C AUTION
- Do not use the product at a voltage exceeding the rated voltage (DC24V +10%-15%).
- Ground the FG terminal of the switching power supply used for the power supply.
Clamp range (rated) | 0.20 mm2 ~ 1.5mm2 | |
Applicable wire |
AWG | AWG24 ~ AWG16 |
Solder plated wire | 0.2 mm2 ~ 1.5mm2 | |
Strand | 0.2 mm2 ~ 1.5mm2 | |
Bar crimp terminal
DIN46228 Part1 |
0.25 mm2 ~ 1.5mm2 | |
Bar crimp terminal (With color)
DIN46228 Part4 |
0.25 mm2 ~ 0.75mm2 | |
Length | 8mm |
Ethernet/IP RS-232C
- When using some weighing devices, such as AD-4212C, the power from this device can be used to operate the weghing device, and wiring of the power supply is not required. Please check the A&D website for the compatible models.
Switch
Change the switch settings according to your operating environment. Be sure to turn off the AD-8552EIP after changing the setting. When the power is turned on, the switch setting change is reflected.
Communication setting for EtherNet/IP
SW-1 | SW-2 | IP address |
0 | 0 | It will be the address set in the web interface. |
0 | 1 | [ 192.168.0.1 ] |
n1 | n2 | [ 192.168.0. (n1×16 + n2×1) ] |
F | E | [ 192.168.0.254 ] |
F | F | [ 192.168.1.10 ] |
Set the IP address so that it does not duplicate that of other devices.
Communication setting for RS-232C
7.2. Communication setting for RS-232C | |||
SW-3 |
No.1 | No.2 | CH1:RS-232C baud rate |
OFF (0) | OFF (0) | 2400 | |
OFF (0) | ON (1) | 9600 | |
ON (1) | OFF (0) | 19200 | |
ON (1) | ON (1) | 38400 | |
No.3 | No.4 | CH2:RS-232C baud rate | |
OFF (0) | OFF (0) | 2400 | |
OFF (0) | ON (1) | 9600 | |
ON (1) | OFF (0) | 19200 | |
ON (1) | ON (1) | 38400 | |
SW-4 |
No.1 | No.2 | CH3:RS-232C baud rate |
OFF (0) | OFF (0) | 2400 | |
OFF (0) | ON (1) | 9600 | |
ON (1) | OFF (0) | 19200 | |
ON (1) | ON (1) | 38400 | |
No.3 | No.4 | CH4:RS-232C baud rate | |
OFF (0) | OFF (0) | 2400 | |
OFF (0) | ON (1) | 9600 | |
ON (1) | OFF (0) | 19200 | |
ON (1) | ON (1) | 38400 | |
7.3. Other settings |
Other settings
Setting of decimal point position
SW-5 | Value stored in AD-8552EIP
(Eg. Weighing output is123.456 g) |
|||
No.2 | No.3 | No.4 | Decimal position | Weighing value |
OFF (0) | OFF (0) | OFF (0) | 0 | 123 |
OFF (0) | OFF (0) | ON (1) | 1 | 1234 |
OFF (0) | ON (1) | OFF (0) | 2 | 12345 |
OFF (0) | ON (1) | ON (1) | 3 | 123456 |
ON (1) | OFF (0) | OFF (0) | 4 | 1234560 |
ON (1) | OFF (0) | ON (1) | 5 | 12345600 |
ON (1) | ON (1) | OFF (0) | 6 | 123456000 |
ON (1) | ON (1) | ON (1) | 3 (AUTO※) | 123456 |
The decimal point value is automatically set according to the weighing value input.
Confirmation
Supply power to the AD-8552EIP to start communication.
It is possible to confirm that the wiring of each cable is connected by the LEDs of the AD-8552EIP
POWER 〇〇 ERR
MS 〇〇 NS
Communication protocol
Please check the A&D website for details on the communication protocol.