LUTRON CL-2006 Free and Total CHLORINE METER User Manual

Home » Lutron » LUTRON CL-2006 Free and Total CHLORINE METER User Manual

LUTRO-logo

LUTRON CL-2006 Free and Total CHLORINE METER

LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-PRODUCT

CHLORINE METER
Model: CL-2006

Your purchase of this Free and Total cholesterol meter marks a step forward for you into the field of precision measurement. Although this CHLORINE METER is a complex and delicate instrument, its durable structure developed. Please read the following instructions carefully and always keep this manual within easy reach.

FEATURES

  • The meter measures the Free and Total chlorine ( CL ) in the 0.00 to 3.50 ppm ( mg/L ).
  • The measuring method is an adaptation of the USEPA Method 330.5 for wastewater and Standard Method 4500-Cl G for drinking water.
  • The advanced optical system is based on a special narrow-band LED lamp that allows the most accurate and repeatable reading.
  • Friendly and powerful calibration functions are able to validate the good performance of your meter at any time.
  • 1.00 ppm Free standard solution and 1.00 ppm Total standard solution are included as the standard accessories.
  • The unique optics structure enables the instrument to read with high resolution: 0.01 ppm ( mg/L ).
  • Splash waterproof on the front panel.
  • Jumbo LCD, easy readout.
  • Microprocessor circuit assures maximum possible accuracy and provides special functions and features.
  • Battery operated for field and on-site testing convenience.
  • Data hold function for freezing the desired value on display.
  • Records Maximum and Minimum reading with Recall.
  • Heavy duty & compact housing with hard carrying case, designed for easy carryout & operation.
  • Auto shut-off is available to save battery life.
  • Application: Test swimming pool, municipal water, food and beverage water, or other aqueous solution where fluid clarity is important.

SPECIFICATIONS

Circuit Custom one-chip of microprocessor LSI circuit.
Display LCD size : 41 mm x 34 mm
Range Free chlorine ( CL ) :

0.00 to 3.50 ppm ( mg/L ).

Total chlorine ( CL ) :

0.00 to 3.50 ppm ( mg/L ).

Resolution 0.01 ppm ( mg/L ).
Accuracy ± 0.02 ppm ( mg/L ).

@ 1.00 ppm ( mg/L )

Light source LED, 525 nm.
Light detector Photodiode
Method The measuring method is an adaptation of the USEPA Method 330.5 and Standard Method 4500-Cl G.

* The reaction between free ( total )

chlorine and the DPD reagent cause a pink tint in the sample.

Response time Less than 10 seconds.
Sample volume 10 mL.
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
Display Sampling Time Approx. 1 second.
Power off Auto shut-off saves battery life or manual off by pushing the button.
Calibration points Zero chlorine.

1.00 ppm ( Free chlorine ).

1.00 ppm ( Total chlorine ).

Operating Temperature 0 to 50 ℃.
Operating Humidity Less than 85% R.H.
Power Supply DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent.
Power Current Stand by Approx. DC 4 mA.
Testing Approx. DC 12 mA.
Weight 320 g/0.70 LB.    @ Battery is included.
Dimension 155 x 76 x 62 mm

( 6.1 x 3.0 x 2.4 inch)

Accessories Included *  Instruction manual…………………….. 1 PC

*  1.0 ppm Free Chlorine standard

solution, CF-01…………………………. 1 PC

*  1.0 ppm Total Chlorine standard solution, CT-01…………………………. 1 PC

*  Zero Chlorine standard solution,

CL-01                                              1 PC

*  Empty testing bottle…………………… 2 PCs

*  Clean cloth………………………………. 1 PC

*  Free Chlorine DPD powder…………… 10 PCs

*  Total Chlorine DPD powder………….. 10 PCs

Optional Accessories *  Free Chlorine DPD powder ( 10 PCs ), Model: CFP-10

*  Total Chlorine DPD powder ( 10 PCs ), Model: CTP-10

*  Empty testing bottle, Model: 0601

*  1.0 ppm Free Chlorine standard solution, Model: CF-01

*  1/0 ppm Total Chlorine standard solution, Model: CT-01

*  Zero Chlorine standard solution,

Model: CL-01

FRONT PANEL DESCRIPTION

LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-1 LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-2

  • 3-1 Cover of Testing bottle
  • 3-2 Container of Testing bottle
  • 3-3 Display
  • 3-4 Hold Button ( Esc Button )
  • 3-5 TEST/CAL Button
  • 3-6 Power Button
  • 3-7 ZERO Button
  • 3-8 REC Button ( MAX, MIN Button )
  • 3-9 Battery Compartment/Cover
  • 3-10 Free Chlorine 1.0 ppm standard solution.
  • 3-11 Total Chlorine 1.0 ppm standard solution
  • 3-12 Zero Chlorine standard solution
  • 3-13 Empty testing bottle 1
  • 3-14 Empty testing bottle 2
  • 3-15 Clean Cloth
  • 3-16 Free Chlorine DPD powder ( 10 PCs )
  • 3-17 Total Chlorine DPD powder ( 10 PCs )

MEASURING PROCEDURE

Measurement ConsiderationLUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-3

  1. There is a ” White mark ” on the edge of the ” Container ” ( 3-2, Fig. 1 ) and also on the up position of ” Testing Bottle ” ( 3-10, 3-11, 3-12 3- 13, 3-14 Fig. 1 ), refer Fig. 2.
  2. When making the measurement ( or calibration ), it should keep the ” Container white mark ” face to face together with the ” Testing bottle white mark “LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-4
  3. insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
  4. Before the measurement, it should be enveloped in the ” Cover ” ( 3-1, Fig. 1 ) completely. Remark: Before the measurement, it should

Measurement
Before the measurement, select the measurement model to ” Free Chlorine ” or ” Total Chlorine ” First, the setting

ZERO setting for the liquid

  1. Power ON the meter by pressing the ” Power Button ” ( 3-6, Fig. 1 ) once.
    The Display ( 3-3, Fig. 1 ) will show the text ” Free ( TOTAL ) ” approx. 1 second. then show ” CAL 0 “, now the meter for the ZERO setting procedures.
  2. Fill the measurement liquid into the ” Testing bottle ” ( 3-11, 3-12, Fig. 1 ) .
  3. LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-5 Insert the ” Testing bottle ” to the bottom of ” Container ” ( 3-2, Fig. 1 ) completely.
    Before the measurement, it should be enveloped in the ” Cover ” ( 3-1, Fig. 1 ) completely.
  4. Press the ” Zero Button ” ( 3-7, Fig. 1 ) once, the display will show the text ” tESt ” ( TEST )” with flashing, then show ” 0.00 “.
    Liquid with the DPD powder
  5. Take the ” Testing bottle ” with the measurement liquid away from the ” Container ” ( 3-2, Fig. 1 ).
    Fill the DPD powder into the ” Testing bottle ” with the measurement liquid. It should shake the at least 10 seconds to keep the solution under uniform conditions.LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-6 Remark :
    1. For the Free Chlorine measurement, it should use the ” Free Chlorine DPD powder “.
    2. For the Total Chlorine measurement, it should use the ” Total Chlorine DPD powder “.
  6. Wait one minute approximately, then insert the ” Testing bottle ” to the bottom of the ” Container ” ( 3-2, Fig. 1 ) completely.
    Before the measurement, it should be enveloped in the
    ” Cover ” ( 3-1, Fig. 1 ) completely.
  7.  Press the” Test Button ” ( 3-5, Fig. 1 ) once, the display will show the text ” tESt ” ( TEST )” with flashing, wait approx. 10 seconds, the Display will show the measurement value.

Wash ( Reins ) the Testing bottle
* After the testing, it should wash ( reins ) the Testing Bottle by the Distill Water

Free/Total Chlorine mode selection

  1. The meter is under the power-off conditions.
    3-8, Fig. 1 ) at the same time ( not release ), then press the ” Hold Button ” ( 3-4, Fig. 1 ) ” REC Button ” ( press the ” Power Button ” ( 3-6, Fig. 1 ) will enter the selection mode to select the function of ” Free Chlorine ” or the ” Total Chlorine “.
  2. Press the ” TEST Button ” ( 3-5, Fig. 1 ) to select the desiring function ( Free Chlorine or Total Chlorine ), then the ” REC Button ” to confirm and save the selection mode into the memory.
    Remark :
    * For the swinging pool application, typical to select the ” Free Chlorine ” function.
    * For the industrial water pollution application, typical to select the ” Total Chlorine ” function.

Data Hold
During the measurement, press the ” Hold Button ” ( 3-4, Fig. 1 ) once will hold the me

  • * Press the ” Hold Button ” once again will release the data hold function.

Data Record ( Max., Min. reading )

  • The data record function records the maximum and minimum readings. Press the ” REC Button ” ( 3-8, Fig. 1 ) once to start the Data Record function and there will be a ” REC. ” symbol on the display.
  •  With the ” REC. ” symbol on the display :
    • a) Press the ” REC Button ” ( 3-8, Fig. 1 ) once, the ” REC. MAX. ” symbol along with the maximum value will appear on the display.
    • b) If intend to delete the maximum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.
      Press the ” REC Button ” ( 3-8, Fig. 1 ) again, the
  • REC. MIN. ” symbol along with the minimum value will appear on the display. If intend to delete the minimum value, just press the ” Hold Button ” ( 3-4, Fig. 1 ) once, then the display will show the ” REC. ” symbol only & execute the memory function continuously.

CALIBRATION PROCEDURE

  1. The meter can be calibrated under the following calibration points:
    Free chlorine
    * Zero
    * 1.00 ppm
    Total chlorine
    * Zero
    * 1.00 ppm
  2. The meter ship along with
    * Zero Chlorine standard solution ( CL-01 ) X 1 PC,
    * 1.0 ppm Free Chlorine standard solution ( CF-01 ) X 1 PC.
    * 1.0 ppm Total Chlorine standard solution ( CT-01) X 1 PC. as the standard accessories.
  3. The complete calibration should be executed by following two solutions:
    * Zero Chlorine standard solution
    * 1.0 ppm standard solution ( Free or Total ).
  4. Zero chlorine calibration
    * Insert the ” Zero Chlorine standard solution into the ” Container ” ( 3-2, Fig. 1) and envelope it in the ” Cover ” ( 3-1, Fig. 1 ) completely, other procedures please refer to Chapter 4-1, 4-2.
    * Press the ” CAL Button ” ( 3-5, Fig. 1 ) continuously until the Display shows the text ” CAL ” and then releases the button.LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-7* During the display show text ” CAL “, press the ” CAL Button ” ( 3-5, Fig. 1 ), the display will showLUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-8 Now the meter is ready for the ” Zero Chlorine ” calibration
    Press ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show the following text with flashing ( approx. 10 seconds ).LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-9 Then the Display will show:LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-10 Now the meter is finished the ” Zero Chlorine ” calibration procedures and is ready for ” 1.00 ppm ” calibration procedures.
  5. 1.00 ppm calibration
    Refer to page 12, when finish the ” Zero Chlorine ” calibration procedures and the Display showLUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-12The  meter is ready for ” 1.0 ppm ” calibration.
    Insert the ” 1.0 ppm standard solution into the ” Container ” ( 3-2 , Fig. 1) and envelope in the
    ” Cover ” ( 3-1, Fig. 1 ) completely, other procedures please refer to chapters 4-1, 4-2.
    Press the ” CAL Button ” ( 3-5, Fig. 1 ) once, the Display will show the following text with flashing ( approx. 10 seconds ).LUTRON-CL-2006-Free-and-Total-CHLORINE-METER-FIG-12 Then the LCD display will return to the normal measurement screen, now the meter is finished the calibration procedures ( Zero, 1.0 ppm calibration ) completely and ready for the measurement.

Remark:
The calibration procedures for the function of ” Total chlorine ” and the ” Free chlorine ” are independent.
The complete calibration procedures for the function of ” Free chlorine ” measurement should be executed by

  • Zero Chlorine standard solution
  • 1.0 ppm Total standard solution

BATTERY REPLACEMENT

  1. When the left corner of the LCD display shows” “, it is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
  2. Loss the ” Battery Cover Screws ” and slide the ” Battery Cover ” ( 3-9, Fig. 1 ) away from the instrument and remove the battery.
  3. Replace with DC 1.5 V battery ( UM4, AAA,
    Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
  4. Make sure the battery cover is secured after changing the battery.